Serviços Personalizados
Journal
Artigo
Indicadores
- Citado por SciELO
- Acessos
Links relacionados
- Similares em SciELO
Compartilhar
Revista Tecnología en Marcha
versão On-line ISSN 0379-3982versão impressa ISSN 0379-3982
Resumo
MUNOZ-PEREZ, Johanna. Optimization of the operating point of an axial impeller without central shaft in blood flow. Tecnología en Marcha [online]. 2022, vol.35, n.3, pp.48-59. ISSN 0379-3982. http://dx.doi.org/10.18845/tm.v35i3.5625.
This article covers the obtention of the mathematical models for the volumetric strain, pressure difference, and flow responses of an axial impeller designed for a ventricular assist device. This impeller has a novel geometry that requires the validation of its behavior at the structural and fluid dynamics level. Based on the specifications required for the application, the optimal operating point of the impeller is obtained in terms of its rotational velocity, material, and blood flow output velocity. Likewise, the range of values of the factors that meet the specifications is obtained. Among the materials included are Platinum and the biocompatible polymer Peek. The analysis is carried out by means of a full factorial design of experiments, with results previously obtained using simulation tools. Obtaining the mathematical models is carried out by debugging the input factors and their interactions.
Palavras-chave : Design of experiments; superimposed contour plot; interaction plot; axial impeller; response optimization.